Search

Henan Tianchi Instrument Equipment Co., Ltd.
Home About Us Products Selling Leads Buying Leads Contact Us
New Product
 
Product Name: Ukraine automated liquid nitrogen storage KGSQ nitrogen tank

Product Details
Brief DescriptionThe choice of Dewar bottle is related to the duration of liquid nitrogen storage and the safety of the sample storage.
Model NoYDZ-500
MaterialStainless steel
Colorwhite
Place of OriginChina
BrandKGSQ
TypeStorage or Transport
applicationfrozen semen
FeaturesVacuum insulation
Special FeaturesThe choice of Dewar bottle is related to the duration of liquid nitrogen storage and the safety of the sample storage.
Payment and Shipping Terms
Fob Price  -    USD
PortQingdao
Payment TermsL/C,D/A,D/P,T/T
Min. Order Quantity1
UnitPiece / Pieces
Capacity6000 sets per month
Delivery Time7-15days

Detailed Product Description
 
Many users don’t understand how to operate a self-pressurized liquid nitrogen tank.Let’s explain it in detail today.
1.Pressure replenishment sequence: close the vent valve,open the boost valve,and open the liquid outlet valve.The self-pressurizing liquid nitrogen tank needs to be pressurized first,and the vent valve must be closed before pressurizing;then open the pressurizing valve to allow the nitrogen in the pressurizing tube to enter the inner tank,and then start the exchange cycle of nitrogen and liquid nitrogen until the pressure increases.to ≥0.05MPa.
After the infusion is completed,first stop the pressurization,then stop draining the liquid nitrogen,and then open the pressure relief exhaust valve,that is,the vent valve.
2.Let it stand for storage and open the vent valve.
During static storage,although there is no pressure,the liquid nitrogen naturally volatilizes,and an outlet is required to release the nitrogen.The vent valve just acts as this "air outlet".

  Products Showcase  
Liquid nitrogen tank (50)
 
Email this page Bookmark this page Print this page